Volume, size and weight
-
Nominal Internal Volume150L
-
Inside DimensionsW500 mm×D500 mm×H600 mm
-
Overall DimensionW750 mm×D1500 mm×H1750 mm (Without Bulge)
-
WeightAbout 260kg
-
Performance
-
Testing Methods
GB/T 5170.2-2008 Temperature Test Equipment
GB/T 5170.5-2008 Damp and Heat Test Equipment
-
Temperature Range-20℃~150℃
-
Temperature Fluctuation≤1℃(No Load and Stable Temperature)
-
Temperature Deviation±2.0℃ (No Load and Stable Temperature)
-
Heating Time-20℃→+150℃ ≤60 min (No load, average nonlinearity)
-
Cooling Time+20℃→-20℃ ≤45 min (No load, average nonlinearity)
-
Humidity Controllable Range Diagram (No Active Wet and Thermal Load)
-
Relative Humidity Deviation±3.0%RH (Humidity>75%RH); ±5.0%RH (Humidity≤75%RH)
-
Meet Testing Method
GB/T 2423.1-2008 Testing Method of Low Temperature Ab
GB/T 2423.2-2008 Testing Method of High Temperature Bb
GJB 150.3A-2009 High Temperature Test
GJB 150.4A-2009 Low Temperature Test
GB/T 2423.3-2006 Constant Damp and Heat Testing Method Cab
GB/T 2423.4-2008 Alternating Damp and Heat Test Method Db
GJB 150.9A-2009 Damp and Heat Testing
GB/T 10586-2006 Technical Conditions for Damp and Heat Test Chamber
GB/T 10592-2008 Technical conditions of high and low temperature test chamber
(Load is not greater than 35kg / m3 steel heat capacity. There is no active wet and heat load during damp and heat testing.)
-
Structure Characteristics
-
Thermal Insulation Enclosure Structure
Outer wall material: High quality cold rolled steel plate with plastic spraying and paint baking treatment on the surface.
Inner wall material: Stainless steel plate SUS304.
Box insulation material: rigid polyurethane foam and glass wool (insulation thickness: 100mm)
Door insulation material: Glass wool.
-
Air Conditioning ChannelCentrifugal fan, heater, evaporator (also dehumidifier), humidifier, etc.
-
Standard Configuration of Test Chamber
Lead hole (multiplicable): φ100mm with 2 lead hole.
(Equipped with soft rubber plugs. One on the left side and the other one on the right side of the box.)
Casters: 4个(With adjusting foot.)
Observation Window: Multi layer hollow electric heating film heating anti fog observation window (located on the door).
Visual Range: About 230×270 mm (wide*high). The glass is equipped with electric mist removal, which can provide the best observation sight line.
Floodlight: 1;
Sample rack (can be added): Two layers of high temperature resistant electrical insulation sample rack. Bearing capacity (uniform distribution): 10kg / layer (the cumulative total bearing capacity of samples in the box shall not exceed 40kg).
-
Door
Single hinged door (left hinge, right handle), equipped with observation window and lighting.
Window frame / door frame anti condensation electric heating device. Double layer silicone rubber sealing strip.
-
Control PanelController display screen, overtemperature protection setter, etc.
-
Refrigeration Unit Room
Refrigeration Unit, Water Tray and Drainage Hole;
Condensing Fan, Lifting Water Pump and Water Storage Tank.
-
Power Distribution Control Cabinet
Main Power Leakage Circuit Breaker (One), Distribution Board (One), Exhaust Fan (One), Ethernet Physical Interface (One);
Temperature and Humidity Controller, AC Contactor, Circuit Breaker, Thermal Relay
Temperature limiting protector, solid state relay and transformer, etc.
-
Heater
Ni Cr alloy electric heating wire heater;
Heater control mode: contactless periodic pulse width modulation, SSR (solid state relay).
-
Humidifier
Stainless steel armored heater;
Heater control mode: contactless periodic pulse width modulation, SSR (solid state relay);
Water level control device, heater anti dry burning device.
-
Refrigeration System
-
Operation ModeMechanical compression single-stage refrigeration mode.
-
Refrigeration Compressor"Taikang" fully enclosed compressor imported from France
-
Main Refrigeration ComponentsExpansion valve, pressure controller, drying filter, refrigeration solenoid valve, liquid reservoir, oil separator, etc.
-
EvaporatorFinned tube heat exchanger (also used as dehumidifier).
-
CondenserAir cooled type: finned tube heat exchanger.
-
Throttling DeviceExpansion Valve / Capillary
-
Refrigerator Control ModeControl system automatically adjusts the operating conditions of the refrigeration unit compressor return air cooling circuit according to the test conditions.
-
RefrigerantR404A (Ozone Wastage Index is 0)
-
Welding TechnologyNitrogen filled in order to protect welding.
-
Control System
-
Controller ModelProfessional Temperature and Humidity Controller
-
MonitorHD color LCD touch screen
-
Operation ModeProgram Mode and Setting Mode
-
Setting ModeColor, Touch, Man-machine Interactive, Chinese/ English Interface
-
Control Mode
Anti integral saturation PID;
BTHC balance temperature and humidity control mode.
-
Temperature and Humidity Measurement MethodClass A armored PT100 sensor
-
Display Accuracy
Temperature: 0.01℃; Humidity: 0.1%RH; Time: 1min.
-
Overtemperature ProtectionIndependent overtemperature protector will protect the shutdown and send an alarm signal when the working room temperature exceeds the temperature set by the protection device.
-
Water Supply System
-
Water Supply ModeWater Pump Lifting
-
Location of Water Supply UnitFront water tank, drawer type water filling.
-
Water Supply Quality RequirementsResistivity≥500Ω·m
-
safety Protection Unit
-
Refrigeration SystemCompressor Overheating, Compressor Overload, Compressor Overpressure, Condensate Fan Overheating
-
Humidification System
Humidification heat pipe dry burning, abnormal water supply and drainage.
-
Test ChamberAdjustable overtemperature, blower motor overload and overheating.
-
Others
Main power supply phase sequence and phase loss protection, leakage protection, overload and short circuit protection, power failure recovery protection.
-
Other Configurations
-
Power CableOne Five Core Cable (Three-Phase Four Wire + Protective Ground Wire)
-
Main Power Leakage Circuit BreakerThree phase four wire and protective ground wire.