諮詢
諮詢
薛瓊
銷售經理 - 薛瓊
羅茹玲
銷售經理 - 羅茹玲
吴燕红
銷售經理 - 吴燕红
葉樹霞
銷售經理 - 葉樹霞
彭君
銷售經理 - 彭君
石璐
銷售經理 - 石璐
王瑤
銷售經理 - 王瑤
蘇曉紅
銷售經理 - 蘇曉紅
王雨薇
銷售經理 - 王雨薇
吳美容
銷售經理 - 吳美容
姚益涵
銷售經理 - 姚益涵
刘芬
銷售經理 - 刘芬
邊群麗
銷售經理 - 邊群麗
楊琪
銷售經理 - 楊琪
江峰
銷售經理 - 江峰
江峰
銷售經理 - 江峰
Select location or language
High Temperature Environmental Chamber
MGW Series (Standard)
  • Volume, size and weight
  • Nominal Internal Volume
    150L
  • Inside Dimensions
    W500 mm×D500 mm×H600 mm
  • Overall Dimension
    W750 mm×D1500 mm×H1750 mm (Without Bulge)
  • Weight
    About 260kg
  • Performance
  • Testing Methods

    GB/T 5170.2-2008 Temperature Test Equipment

    GB/T 5170.5-2008 Damp and Heat Test Equipment

  • Temperature Range
    -20℃~150℃
  • Temperature Fluctuation
    ≤1℃(No Load and Stable Temperature)
  • Temperature Deviation
    ±2.0℃ (No Load and Stable Temperature)
  • Heating Time
    -20℃→+150℃ ≤60 min (No load, average nonlinearity)
  • Cooling Time
    +20℃→-20℃ ≤45 min (No load, average nonlinearity)
  • Humidity Controllable Range Diagram (No Active Wet and Thermal Load)
  • Relative Humidity Deviation
    ±3.0%RH (Humidity>75%RH); ±5.0%RH (Humidity≤75%RH)
  • Meet Testing Method

    GB/T 2423.1-2008   Testing Method of Low Temperature Ab

    GB/T 2423.2-2008   Testing Method of High Temperature Bb

    GJB 150.3A-2009    High Temperature Test

    GJB 150.4A-2009    Low Temperature Test

    GB/T 2423.3-2006   Constant Damp and Heat Testing Method Cab

    GB/T 2423.4-2008   Alternating Damp and Heat Test Method Db

    GJB 150.9A-2009    Damp and Heat Testing

    GB/T 10586-2006   Technical Conditions for Damp and Heat Test Chamber

    GB/T 10592-2008   Technical conditions of high and low temperature test chamber

    (Load is not greater than 35kg / m3 steel heat capacity. There is no active wet and heat load during damp and heat testing.)

  • Structure Characteristics
  • Thermal Insulation Enclosure Structure

    Outer wall material: High quality cold rolled steel plate with plastic spraying and paint baking treatment on the surface.

    Inner wall material: Stainless steel plate SUS304.

    Box insulation material: rigid polyurethane foam and glass wool (insulation thickness: 100mm)

    Door insulation material: Glass wool.

  • Air Conditioning Channel
    Centrifugal fan, heater, evaporator (also dehumidifier), humidifier, etc.
  • Standard Configuration of Test Chamber

    Lead hole (multiplicable): φ100mm with 2 lead hole.

    (Equipped with soft rubber plugs. One on the left side and the other one on the right side of the box.)

    Casters: 4个(With adjusting foot.)


    Observation Window: Multi layer hollow electric heating film heating anti fog observation window (located on the door).

    Visual Range: About 230×270 mm (wide*high). The glass is equipped with electric mist removal, which can provide the best observation sight line.

    Floodlight: 1;

    Sample rack (can be added): Two layers of high temperature resistant electrical insulation sample rack. Bearing capacity (uniform distribution): 10kg / layer (the cumulative total bearing capacity of samples in the box shall not exceed 40kg).

  • Door

    Single hinged door (left hinge, right handle), equipped with observation window and lighting.

    Window frame / door frame anti condensation electric heating device. Double layer silicone rubber sealing strip.

  • Control Panel
    Controller display screen, overtemperature protection setter, etc.
  • Refrigeration Unit Room

    Refrigeration Unit, Water Tray and Drainage Hole;

    Condensing Fan, Lifting Water Pump and Water Storage Tank.


  • Power Distribution Control Cabinet

    Main Power Leakage Circuit Breaker (One), Distribution Board (One), Exhaust Fan (One), Ethernet Physical Interface (One);

    Temperature and Humidity Controller, AC Contactor, Circuit Breaker, Thermal Relay

    Temperature limiting protector, solid state relay and transformer, etc.

  • Heater

    Ni Cr alloy electric heating wire heater;

    Heater control mode: contactless periodic pulse width modulation, SSR (solid state relay).


  • Humidifier

    Stainless steel armored heater;

    Heater control mode: contactless periodic pulse width modulation, SSR (solid state relay);

    Water level control device, heater anti dry burning device.


  • Refrigeration System
  • Operation Mode
    Mechanical compression single-stage refrigeration mode.
  • Refrigeration Compressor
    "Taikang" fully enclosed compressor imported from France
  • Main Refrigeration Components
    Expansion valve, pressure controller, drying filter, refrigeration solenoid valve, liquid reservoir, oil separator, etc.
  • Evaporator
    Finned tube heat exchanger (also used as dehumidifier).
  • Condenser
    Air cooled type: finned tube heat exchanger.
  • Throttling Device
    Expansion Valve / Capillary
  • Refrigerator Control Mode
    Control system automatically adjusts the operating conditions of the refrigeration unit compressor return air cooling circuit according to the test conditions.
  • Refrigerant
    R404A (Ozone Wastage Index is 0)
  • Welding Technology
    Nitrogen filled in order to protect welding.
  • Control System
  • Controller Model
    Professional Temperature and Humidity Controller
  • Monitor
    HD color LCD touch screen
  • Operation Mode
    Program Mode and Setting Mode
  • Setting Mode
    Color, Touch, Man-machine Interactive, Chinese/ English Interface
  • Control Mode

    Anti integral saturation PID;

    BTHC balance temperature and humidity control mode.


  • Temperature and Humidity Measurement Method
    Class A armored PT100 sensor
  • Display Accuracy

    Temperature: 0.01℃; Humidity: 0.1%RH; Time: 1min.

  • Overtemperature Protection
    Independent overtemperature protector will protect the shutdown and send an alarm signal when the working room temperature exceeds the temperature set by the protection device.
  • Water Supply System
  • Water Supply Mode
    Water Pump Lifting
  • Location of Water Supply Unit
    Front water tank, drawer type water filling.
  • Water Supply Quality Requirements
    Resistivity≥500Ω·m
  • safety Protection Unit
  • Refrigeration System
    Compressor Overheating, Compressor Overload, Compressor Overpressure, Condensate Fan Overheating
  • Humidification System

    Humidification heat pipe dry burning, abnormal water supply and drainage.

  • Test Chamber
    Adjustable overtemperature, blower motor overload and overheating.
  • Others

    Main power supply phase sequence and phase loss protection, leakage protection, overload and short circuit protection, power failure recovery protection.

  • Other Configurations
  • Power Cable
    One Five Core Cable (Three-Phase Four Wire + Protective Ground Wire)
  • Main Power Leakage Circuit Breaker
    Three phase four wire and protective ground wire.
Neware and Cookies
This site uses cookies. By continuing to browse the site you are agreeing to our use of cookies.Read our private policy